LCD Interface Block

The interface to the LCD is via the LCD JILI interface, designed to provide the following:
-
-
-
-
5-4-3-3

CardBus Interface

The PCMCIA slot supports a single type-II or a single type III Card.
The CardBus controller and Power Interface switch supports hot-plug capability.
5-4-4

Bluetooth Module and Antenna

The Bluetooth Module and Antenna enable the Voluson i / Voluson e portable ultrasound system to
operate in a wireless LAN network, which enables the system to connect to external applications, to
save or retrieve data from other network clients.
The Bluetooth module is provided through the upper USB port, using an USB-to-Bluetooth adaptor.
5-4-5

Wireless Network Adapter

The Voluson i / Voluson e has wireless capabilities using the Cisco Aironet Wireless LAN Adapter via
the PCMCIA port.
5-4-6

Wireless Network Interface

The Voluson i / Voluson e supports a Wireless Network Interface based on industry standards to provide
easy-to-use and compatible high-speed wireless connectivity. For details regarding installation, refer to
Section 3-5-4 "Connecting the Wireless Network Adapter/Interface" on page
The Wireless Network Interface provides a mobile network connection to the local area network.
10
EVISION
A generic interface to LCD panels
Backlight illumination control
Backlight ON/OFF control
Supports LVDS interface
Chapter 5 - Components and Functions (Theory)
V
i / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
3-43.
5-23
e

Advertisement

Table of Contents
loading

Table of Contents