Networking Setup Requirements; Stand-Alone Unit (Without Network Connection); Unit Connected To Hospital's Network; Purpose Of The Dicom® Network Function - GE Voluson i BT06 Service Manual

Table of Contents

Advertisement

GE
D
KTI106052, R
IRECTION
2-3-4

Networking Setup Requirements

2-3-4-1

Stand-alone Unit (without Network Connection)

None
2-3-4-2

Unit Connected to Hospital's Network

Supported networks:
Ethernet
Wireless LAN
2-3-4-3
Purpose of the DICOM® Network Function
DICOM® (Digital Imaging and Communications in Medicine) services provide the operator with
clinically useful features for moving images and patient information over a hospital network.
Examples of DICOM® services include the transfer of images to workstations for viewing or transferring
images to remote printers. As an added benefit, transferring images in this manner frees up the on-
board monitor and peripherals, enabling viewing to be done while scanning continues.
With DICOM®, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-3-4-4
DICOM® Option Pre-installation Requirements
To configure the Voluson i / Voluson e ultrasound unit to work with other network connections, the
network administrator must provide some necessary information.
Use the
• Voluson i / Voluson e Details:
• Routing Information:
• DICOM® Application
Information:
Installation see:
2-8
10
EVISION
Connectivity Setup Worksheet on page 3-119
Section 3-12 "Network IP Address Configuration" on page
Section 2-3 - Facility Needs
to record required information that must include:
DICOM® network details for the Voluson i / Voluson e unit, incl.
the host name, local port, IP address, AE title and net mask.
IP addresses for the default gateway and other routers in use at
the site.
Details of DICOM® devices in use at the site, including the
DICOM® host name, AE title, DICOM® port number and IP
addresses.
V
i / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
3-116.