Basic Probe Cleaning and/or Disinfection

Refer to the Basic User Manual of Voluson i / Voluson e for details on cleaning and desinfection.
To help protect yourself from blood borne diseases, wear approved disposable gloves.
WARNING
WARNING
These are made of nitrile derived from vegetable starch to prevent allergic latex
reactions.
Failure to follow the prescribed cleaning or disinfection procedures will void the probe's
CAUTION
warranty.
!! CAUTION:
DO NOT soak or wipe the lens with any product not listed in the Voluson i / Voluson e Basic User
Manual and/or care card (supplied with the probe). Doing so could result in irreparable damage
to the probe and/or Voluson i / Voluson e system.
Follow the Care Card instructions supplied with each probe (inside the transducer boxes)
CAUTION
for disinfectants and gels that are compatible with the surface material of the probes.
!! CAUTION:
For the latest list of compatible cleaning solutions and disinfectants refer to:
NOTICE
http://www.gehealthcare.com/usen/ultrasound/products/probe_care.html
Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.
CAUTION
!! CAUTION:
CAUTION
CAUTION:
Please be aware of the
sensitive probe head.
TAKE EXTREME CARE!
10-10
10
EVISION
NEVER place or store a
probe on its scan head!
Section 10-5 - System Maintenance
V
i / V
OLUSON
S
ERVICE
When disinfecting a probe,
ensure that there is sufficient
space between the probe and
the container bottom!
e
OLUSON
M
ANUAL