Battery Performance Maintenance - GE LOGIQ e Basic Service Manual

Hide thumbs Also See for LOGIQ e:
Table of Contents

Advertisement

Battery Performance Maintenance

Battery discharge
No
Replace battery
Figure 10-5. Flow chart of Battery Performance Maintenance
NOTE:
NOTE:
LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Battery replacement every three years is recommended.
It is recommended to do battery performance maintenance one
time per year.
Please follow the flow chart below to carry out battery
performance maintenance.
>45 minutes?
Yes
Battery charge > 3 hour
Battery power > 90%?
Yes
Battery is Ok
Disconnect all probes when discharge battery.
Discharge the battery to let the system automatically shut down.
No
Battery charge > 3 hour
Battery power > 90%?
Yes
Discharge
>45 minutes?
Yes
Battery is Ok
System maintenance
No
Replace battery
No
Replace battery