Starting a case using default settings

Case Defaults contain five case type selections. Each case type has preset values for Ideal Weight, Age, and
Volume Apnea Alarm. The first four default case types are configured and named by the Super User. The fifth
default case is Last Case.
1. Set the Bag/Vent switch to Bag.
2. Select Start Case.
The Case Defaults selection shows the first preset case type
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm show the default settings that correspond to
the case type shown
3. Verify or change the Case Defaults selected.
4. Verify the settings are clinically appropriate.
5. Select Start Case Now. Gas flow starts.
3
4 4
5
23
2