Active alarm touch points

2-14
1.
Touch the measured value to access the Alarm Setup menu.
2.
The Alarm Setup menu displays.
3.
Select the alarm limit and set it to the correct value. Touch the
value on the touchscreen or push the ComWheel to confirm the
desired setting.
4.
Push the Home key, touch the waveform area of the display, or
select Close to close the menu.
When an alarm sounds the alarm message is displayed at the top of
the screen and, if applicable, the alarming numeric field and digit
field flashes. The Alarm messages at the top of the screen are
message alerts only and not active touch points.
1.
Touch the flashing numeric field to access the Alarm Setup
menu and alarm limits for the active alarm.
2.
The Alarm Setup menu displays with the active alarm limit
highlighted. For example: If the 'Ppeak high' alarm activates, the
high alarm limit setting for Ppeak displays with the highlight.
3.
Select the active alarm limit and change it to the desired setting.