R&S NRP18S Series User Manual page 98

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
4. If needed, load uncertainty data. See
on page 98.
5. Check the entries in the "S-Parameter Device Mnemonic", "Lower Power Limit/W"
and "Upper Power Limit/W" fields and change them, if necessary.
For example, the lower and upper power limits are deduced from the power limits
of the sensor itself and the minimum attenuation of the S-parameter device. If the
"Upper Power Limit/W" entry is higher than the power dissipation rating of the
attenuator, reduce it accordingly.
6. Click "OK" to apply the changes.
If you want to discard the change, click "Cancel".
To load an uncertainty parameter file
1. In the device table, double-click an entry. See also
2. Select "File" > "Import uncertainties".
3. Select the file you want to import, and confirm with "Open".
The data from the selected file is loaded in the device table.
User Manual 1178.3686.02 ─ 05
Remote Control Commands
Configuring Basic Measurement Parameters
"To load an uncertainty parameter file"
"Device table"
on page 95.
98

Advertisement

Table of Contents
loading

Table of Contents