R&S NRP18S Series User Manual page 140

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
10.3.5 Device Status Register
Sensor error summary
Reference PLL locked
Figure 10-3: Device status register
Querying the register:
STATus:DEVice:CONDition?
STATus:DEVice[:EVENt]?
Querying the static errors:
SYSTem:SERRor?
Table 10-5: Used device status bits and their meaning
Bit no.
0
1
2
3
4
User Manual 1178.3686.02 ─ 05
0
Sensor error
1
Sensor error
2
3
Sensor error
Sensor error
4
0
5
0
6
7
Legacy locked
8
0
9
10
0
11
0
12
0
0
13
14
0
0
15
Short description
Sum of SERR bits
SERR measurement not possi-
ble
SERR erroneous results
SERR warning
SERR critical
+
Bit is set if
Sum/combination of SERR bits 1 to 4.
Static error exists. The measurement result is possibly incor-
rect.
Static error exists. Status LED of the power sensor is slowly
flashing red.
Critical static error exists. Status LED of the power sensor is
fast flashing red.
Remote Control Basics
Status Reporting System
140

Advertisement

Table of Contents
loading

Table of Contents