R&S NRP18S Series User Manual page 131

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
Example:
Definition: INITiate[:IMMediate]
Command: INIT:IMM is equivalent to INIT
Parameters
Parameters must be separated from the header by a "white space". If several parame-
ters are specified in a command, they are separated by a comma.
For a description of the parameter types, refer to
on page 131.
Special characters
|
[ ]
{ }
10.2.3 SCPI Parameters
Many commands are supplemented by a parameter or a list of parameters. The
parameters must be separated from the header by a "white space" (ASCII code 0 to 9,
11 to 32 decimal, e.g. blank). Allowed parameters are:
Numeric values
Special numeric values
Boolean parameters
Text
Character strings
Block data
The parameters required for each command and the allowed range of values are
specified in the command description.
Numeric values
Numeric values can be entered in any form, i.e. with sign, decimal point and exponent.
Values exceeding the resolution of the instrument are rounded up or down. The man-
tissa can comprise up to 255 characters, the exponent must lie inside the value range
-32000 to 32000. The exponent is introduced by an "E" or "e". Entry of the exponent
alone is not allowed.
User Manual 1178.3686.02 ─ 05
Parameters
A vertical stroke in parameter definitions indicates alternative possibilities in the sense of "or". The
effect of the command differs, depending on which parameter is used.
Mnemonics in square brackets are optional and can be inserted into the header or omitted.
Example: INITiate[:IMMediate]
INIT:IMM is equivalent to INIT
Parameters in curly brackets are optional and can be inserted once or several times, or omitted.
Remote Control Basics
SCPI Command Structure
Chapter 10.2.3, "SCPI
Parameters",
131