Parameters:
<update_time>
SYSTem:SERRor?
Queries the next static error, if available. Static errors are more severe than normal
error conditions, which can be queried using SYSTem:ERRor[:NEXT]?.
Normal errors result from, for example, unknown commands or syntax errors and gen-
erally affect a single parameter or setting. Static errors, as a rule, prevent the execution
of normal measurements.
Static errors occur when you select conflicting settings. This could, for example, occur
in timeslot measurement mode with the following settings:
Width of a timeslot: 100 µs
Exclude time at the start of the slot: 40 µs
Exclude time at the end of the slot: 60 µs
Then there is "nothing left" to be measured, and a static error appears.
Usage:
SYSTem:SERRor:LIST:ALL?
Returns a list of all static errors that have occurred but have already been resolved.
For example, an overload of a short duration.
Example:
Usage:
SYSTem:SERRor:LIST[:NEXT]?
Queries the list of all static errors that have occurred but have already been resolved
for the oldest entry and removes it from the queue. The response consists of an error
number and a short description of the error.
Example:
User Manual 1178.3686.02 ─ 05
Range:
0.0 to 10.0
*RST:
0.1
Default unit: Seconds
Query only
SYST:SERR:LIST:ALL?
Response: 0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification",0,"removed at uptime:2944;
notice; auto-averaging exceeded maximum time;
Notification".
Query only
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Remote Control Commands
Configuring the System
114