R&S NRP18S Series User Manual page 149

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
Sensor waiting for trigger
Figure 10-11: Operation trigger status register
Querying the register:
STATus:OPERation:TRIGger:CONDition?
STATus:OPERation:TRIGger[:SUMMary][:EVENt]?
Table 10-13: Used operation trigger status bits and their meaning
Bit no.
1
10.3.8.4
Operation Sense Status Register
The CONDition register contains information whether a power sensor is being initial-
ized. The EVENt register contains information whether an initialization was started or
completed since the last query.
A power sensor is initialized if:
Supply voltage is switched on (power-up).
Sensor was connected.
Reset was performed using:
User Manual 1178.3686.02 ─ 05
0
0
1
2
0
3
0
4
0
0
5
0
6
0
7
0
8
9
0
10
0
11
0
0
12
13
0
0
14
0
15
Short description
Sensor waiting for trigger
*RST
SYSTem:PRESet
+
Bit is set if
Sensor is waiting for a trigger event. When the trigger event
occurs, the sensor changes into the measuring state.
Remote Control Basics
Status Reporting System
149