R&S
NRP18S-xx
Example: Repeating termination control
Further settings for this example:
[SENSe<Sensor>:]AVERage:TCONtrol
Every chopper phase is started by a trigger event and lasts the defined aperture time.
During a chopper phase, the trigger synchronization is high
ON). After 2 chopper phases, 1 measurement is completed. As defined by the average
count, after 2 measurements, the result is averaged and available.
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Measurement result
6 = Trigger synchronization
7 = Return to the start of the measurement cycle
User Manual 1178.3686.02 ─ 05
2
3
1
REPeat
4
7
Remote Control Commands
Controlling the Measurement
(TRIGger:SYNC:STATe
5
6