R&S NRP18S Series User Manual page 68

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
SENS:FUNC
"XTIMe:POWer"
Parameters:
<mode>
Example:
[SENSe<Sensor>:]AUXiliary <mode>
Enables the measurement of additional measured values that are determined together
with the main measured value.
Parameters:
<mode>
User Manual 1178.3686.02 ─ 05
Possible CALC:FEED
"POWer:RANDom"
"POWer:TRACe"
"POWer:PEAK:TRACe"
"POWer:RANDom:TRACe"
*RST:
"POWer:AVERage"
The following sequence of commands configures a peak trace
measurement:
*RST
SENSe:FUNCtion "XTIMe:POWer"
SENSe:FREQuency 1.0e9
SENSe:TRACe:POINts 500
SENS:TRAC:TIME 20e-3
TRIGger:SOURce INTernal
TRIGger:SLOPe POSitive
TRIGger:DTIMe 0.001
TRIGger:HYSTeresis 0.1
TRIGger:LEVel 30e-6
SENSe:TRACe:AVERage:COUNt 8
SENSe:TRACe:AVERage:STATe ON
CALCulate:FEED "POWer:PEAK:TRACe"
INITiate
FETCh?
NONE | MINMax | RNDMax
NONE
No additional values are measured.
MINMax
Minima and maxima of the trace are transmitted together with
the measured value.
Usually, extreme values are lost due to averaging the measured
values.
RNDMax
Randomly selected samples are transmitted. All evaluations use
these values instead of the average values.
*RST:
NONE
Remote Control Commands
Configuring and Retrieving Results
Meaning
Randomly selected value from the
measurement interval
Measurement sequence
Peak value of the samples per
trace point
Randomly selected value of the
samples per trace point
68