NRP18S-xx
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a measurement with enabled data buffer
mode.
Usage:
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Queries the last valid measurement result.
Usage:
FETCh<Sensor>[:SCALar][:POWer]:BURSt?
Queries the last valid measurement result for the burst average measurement mode.
Usage:
FETCh<Sensor>[:SCALar][:POWer]:TSLot?
Queries the last valid measurement result of the timeslot average measurement mode.
Usage:
CALCulate:FEED <mode>
If you query measurement data using
AVG]?, the power sensor returns data of the measurand that was configured before.
Generally, this measurand is the average power. However, the power sensor can also
output data of other measurands.
To configure which measurand the
command sends, use the CALCulate:FEED command before the measurement is ini-
tiated. Depending on the measurement mode, the following settings are possible:
SENS:FUNC
"POWer:AVG"
"POWer:BURSt:AVG"
"POWer:TSLot:AVG"
User Manual 1178.3686.02 ─ 05
Query only
Query only
Query only
Query only
FETCh<Sensor>[:SCALar][:POWer][:
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Possible CALC:FEED
"POWer:AVERage"
"POWer:PEAK"
"POWer:RANDom"
"POWer:AVERage"
"POWer:PEAK"
"POWer:RANDom"
"POWer:AVERage"
"POWer:PEAK"
Remote Control Commands
Configuring and Retrieving Results
Meaning
Average value
Peak value
Randomly selected value from the
measurement interval
Average value
Peak value
Randomly selected value from the
measurement interval
Average value
Peak value
67