NRP18S-xx
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE <type>
Sets the automatic averaging filter mode.
Parameters:
<type>
[SENSe<Sensor>:]AVERage:TCONtrol <mode>
Defines how the measurement results are output. This is called termination control.
See also
Parameters:
<mode>
Example:
[SENSe<Sensor>:]AVERage[:STATe] <state>
Enables or disables the averaging filter.
Parameters:
<state>
User Manual 1178.3686.02 ─ 05
RESolution | NSRatio
RESolution
The usual mode for the power sensors.
NSRatio
Predefines the compliance to an exactly defined noise compo-
nent.
*RST:
Chapter 8.5, "Controlling the
MOVing | REPeat
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using
on page 81.
COUNt
*RST:
AVER:TCON REP
*RST:
Remote Control Commands
Configuring Basic Measurement Parameters
RESolution
Measurement",
on page 49.
[SENSe<Sensor>:]AVERage:
REPeat
ON
84