Example: Moving termination control
Further settings for this example:
[SENSe<Sensor>:]AVERage:TCONtrol
Every chopper phase is started by a trigger event and lasts the defined aperture time.
During a chopper phase, the trigger synchronization is high
ON). Every measurement provides a result. After 2 measurements, when the average
count is reached, settled data are available.
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Trigger synchronization
6 = Measurement result before average count is reached
7 = Averaged measurement result after average count is reached
8 = Return to the start of the measurement cycle
User Manual 1178.3686.02 ─ 05
2
3
8
1
MOVing
4
6
Remote Control Commands
Controlling the Measurement
(TRIGger:SYNC:STATe
5
7
57