R&S
NRP18S-xx
MT = 2 * AC * APER + (2 * AC - 1) * 100
with:
MT: overall measurement time
AC: average count
APER: aperture time
100 μs is the time for switching the chopper phase.
Using
surement as follows:
●
Chopper is disabled.
●
Average count is set to 1, no matter which average count you have set.
Thus, the overall measurement time is only defined by the aperture time, and the mea-
surement time for a fast measurement is calculated as follows:
MT = APER
The fast measurement setting delivers up to 100 000 measurements per second with-
out any blind time over randomly long time periods. Programming examples are given
in
Chapter 9.2, "Performing the Fastest Measurement in Continuous Average
on page 120.
Remote commands:
[SENSe<Sensor>:][POWer:][AVG:]APERture
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
[SENSe<Sensor>:][POWer:][AVG:]FAST
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe
[SENSe<Sensor>:][POWer:][AVG:]APERture <integration_time>
Sets the duration of the sampling window in the continuous average mode. During this
time interval, the average signal power is measured.
Parameters:
<integration_time>
Example:
User Manual 1178.3686.02 ─ 05
[SENSe<Sensor>:][POWer:][AVG:]FAST
The minimum value is implemented for fast unchopped continu-
ous average measurements. See also
Fast Unchopped Continuous Average
on page 122.
Range:
*RST:
Default unit: Seconds
APER 0.02
Remote Control Commands
Configuring the Measurement Modes
μ s
ON, you can accelerate the mea-
...................................................................... 70
............................................................... 71
............................................................71
.............................................................. 71
.................................................................71
...............................................................71
............................................................................ 72
.......................................................... 72
8.0e-6 to 2.00
0.02
Mode",
Chapter 9.2.2, "Triggered
Measurement",
70