R&S NRP18S Series User Manual page 76

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
Parameters:
<time>
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID[:STATe] <state>
Enables or disables the blanking out of time intervals in the timeslots.
Parameters:
<state>
8.7.4 Trace Measurement
The trace measurement is used to acquire the course of power over a certain time.
During the measurement time ([SENSe<Sensor>:]TRACe:TIME) a large number of
measurements are made and the result is returned to the user as an array of values
with a predefined size [SENSe<Sensor>:]TRACe:POINts. The length of an individ-
ual measurement(-point) is determined from the ratio of measurement time and mea-
surement points. The entire result is called a "trace". Each trace must be triggered sep-
arately.
Remote commands:
[SENSe<Sensor>:]TRACe:AVERage:COUNt
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
[SENSe<Sensor>:]TRACe:DATA?
[SENSe<Sensor>:]TRACe:MPWidth?
[SENSe<Sensor>:]TRACe:OFFSet:TIME
[SENSe<Sensor>:]TRACe:POINts
[SENSe<Sensor>:]TRACe:REALtime
[SENSe<Sensor>:]TRACe:TIME
[SENSe<Sensor>:]TRACe:AVERage:COUNt <count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Averaging is only effective, if
set.
Parameters:
<count>
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol <mode>
Available in trace mode.
Defines how the measurement results are output. This is called termination control.
User Manual 1178.3686.02 ─ 05
Range:
0.00 to 0.10
*RST:
0.00
Default unit: Seconds
*RST:
OFF
...................................................................... 76
...................................................................76
..................................................................... 77
..................................................................................... 77
................................................................................ 80
........................................................................... 80
..................................................................................... 80
.................................................................................80
........................................................................................81
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
Range:
1 to 65536
*RST:
4
Remote Control Commands
Configuring the Measurement Modes
ON is