R&S NRP18S Series User Manual page 74

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
[SENSe<Sensor>:][POWer:]BURSt:LENGth?
Queries the length of a burst (pulse interval), the time between the trigger point of the
measurement and the time the trigger logic detects the end of the pulse (see
ure
8-2).
Usage:
8.7.3 Timeslot Average Measurement
The Timeslot Average mode is used to measure the average power of a definable
number of successive timeslots within a frame structure with equal spacing. The mea-
surement result is an array with the same number of elements as timeslots. Each ele-
ment represents the average power in a particular timeslot.
Fig.8-3
meaning of the parameters that can be configured.
Trigger level
SENS:TSLOT:MID:OFFS:TIME
SENS:TSLOT:MID:TIME
Figure 8-3: Timeslot parameters
Triggering a timeslot average measurement
In the timeslot mode, internal and external trigger events from the signal are evaluated
depending on the settings of the
the
TRIGger:DELay
sured coincides with the delayed trigger point.
Defining a time interval for the measurement
Additionally, time intervals that are to be excluded from the measurement can be set at
the beginning and at the end of the measurement interval, with the commands
User Manual 1178.3686.02 ─ 05
Query only
shows a graphical representation of a timeslot average measurement and the
T'Slot 1
Power
Trigger event
parameter to ensure that the beginning of the first slot to be mea-
Configuring the Measurement Modes
SENS:TSLOT:COUNT 3
T'Slot 2
T'Slot 3
SENS:TSLOT:WIDTH
parameter. It is essential to define
TRIGger:SOURce
Remote Control Commands
Fig-
Time