R&S NRP18S Series User Manual page 32

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
Table 5-6: Offset settings for operating an R&S NRP18S-xx with a different attenuator
Sensor operation mode...
... with a different S-parameter
device not supplied in the
delivery
Setup
Figure 5-4: Setup with an R&S NRP2 base unit
1 = Signal source
2 = Attenuator
3 = R&S NRP18S-xx power sensor
User Manual 1178.3686.02 ─ 05
S-Parameters tool dialogs...
Since the R&S NRP2 does not support selecting one out of several S-
Parameter devices, you must activate the correct device via the S-Param-
eters tool included in the R&S NRP Toolkit, see
the S-Parameters
Tool",
on page 91.
Activate the "S-Parameter Device" in the R&S NRP2 "Offset" dialog.
Operating Concepts
R&S NRP2
Chapter 8.8.4.5, "Using