R&S NRP18S Series User Manual page 141

High-power three-path diode power sensors
Hide thumbs Also See for NRP18S Series:
Table of Contents

Advertisement

®
R&S
NRP18S-xx
Bit no.
7
8
10.3.6 Questionable Status Register
Contains information on questionable sensor states that occur if the power sensor is
not operated in compliance with its specifications.
User Manual 1178.3686.02 ─ 05
Short description
Legacy locked state
Reference PLL locked state
Bit is set if
The power sensor is locked in the NRP legacy mode. Via the
SCPI channels (USBTMC or TCP/IP), only query commands
can be sent, but no setting commands.
When the first setting command is sent, the NRP legacy inter-
face takes precedence over all other command channels. This
bit is set to 1, and all other channels can only execute query
commands. If a setting command is sent via a different chan-
nel, the power sensor indicates an error:
-200,"Execution error; sensor in LEGACY mode"
To leave this operating mode, close the NRP legacy channel.
Either close the application which opened the NRP legacy
channel or close at least the connection to the power sensor.
PLL for the clock reference is synchronized. The bit is useful
when selecting an external clock source.
The following states are possible:
Internal clock
([SENSe<Sensor>:]ROSCillator:
INT) :
SOURce
1 (always)
External clock
([SENSe<Sensor>:]ROSCillator:
EXT) :
SOURce
1 if the sensor was able to synchronize with external
clock
0 if the sensor could not synchronize with external
clock
Remote Control Basics
Status Reporting System
141