R&S
NRP18S-xx
Example: Moving termination control
Further settings for this example:
[SENSe<Sensor>:]AVERage:TCONtrol
TRIGger:COUNt
Every measurement is started by a trigger event. Due to the chopper phases, one
measurement lasts twice the defined aperture time. During each measurement, the
trigger synchronization is high
vides a result. During the settling phase, the amount of the result is already correct, but
the noise is higher. After 4 measurements, when the average count is reached, settled
data are available.
When the trigger count is reached
returns to the idle state.
2
3
1
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Trigger synchronization
6 = Measurement result before average count is reached
7 = Averaged measurement result after average count is reached
8 = Return to idle state after trigger count (= 16 in this example) is reached
8.5.4.2
Trace Mode
General settings for the first two examples:
INITiate:CONTinuous
[SENSe<Sensor>:]AVERage:COUNt
User Manual 1178.3686.02 ─ 05
16
4
6
6
8
MOVing
(TRIGger:SYNC:STATe
(TRIGger:COUNt
6
ON
2
Remote Control Commands
Controlling the Measurement
ON). Every measurement pro-
on page 59), the power sensor
7
7
5
55