Maintenance Schedule:General
Service Manual: Mode
2.0
Maintenance Schedule
2.1
General
All Novametrix Medical Systems Inc. equipment is warrantied for a pe
of one year from the date of shipment, except for certain disposable proc
and products with stated guarantees other than one year. However, to in
the functional integrity of the monitor, Novametrix recommends the follo\
maintenance schedules be adopted with regard to the Model 515.
At the customer's request, Novametrix will provide repair and caiibrt
services under the terms of a Service Contract. Contact the Novametrix
vice Department at 1-800-243-3444 (in Connecticut call Collect 203-:
7701) for contract details.
2.2
Battery Maintenance
The monitor was shipped from the factory with a fully charged bat
Since the monitor draws some battery power even while turned off, it is rec
mended that the monitor be operated on line power for the first 24 hours t(
low ample time for the battery to fully recharge.
In the case of prolonged periods of storage or shipment the battery
discharge to a point where the unit will not turn on. Should this occur, plu
the line cord, set the switch on the rear panel to the I position, turn the moi
off (ensure that the green ~ LED on the front panel is illuminated), and a
the monitor to charge for 24 hours before switching it on.
To insure maximum battery life and performance, the Model 515 shoul
operated in the battery mode until the low battery indicator lights (line
disconnected) at least once each month. The line cord should then be re
nected and the battery given sufficient time to recharge. The monitor ma;
operated normally on line power while the battery is recharging. Approxin
ly 16 hours are required to fully recharge a discharged battery.
Every six months the battery should be visually inspected for defects
eluding but not limited to, swollen or misshapen sides, leaking battery acid,
posed conductors or frayed insulation. If any of these conditions exist, cor
the Novametrix Service Department for a battery replacement. Replacer
batteries are subject to a replacement charge.
2.3
Fuses and Line Cord
Should the fuse(s) need replacement be sure to replace with the same
and rating; 250 V 1/2A slow-blow for lOOV and 120V units; 250 V 1/4A si