3.1 PERFORMANCE SPECIFICATIONS
Systolic Pressure Readout
Number of Digits:
Accuracy*:
Range:
Diastolic Pressure Readout
Number of Digits:
Accuracy*:
Range:
*Tested per ANSI/AAMI SP10-1992 methods.
Accutorr Plus Service Manual
Chapter 3 - Specifications
3
Mean error less than ± 5 mmHg, Standard deviation
less than ± 8 mmHg.
Adult Mode: 55 to 260 mmHg
Pediatric Mode: 55 to 160 mmHg
Neonatal Mode: 45to 120 mmHg
3
Mean error less than ± 5 mmHg, Standard deviation
less than ± 8 mmHg.
Adult Mode: 30 to 200 mmHg
Pediatric Mode: 30 to 150 mmHg
Neonatal Mode: 25 to 100 mmHg
3-1
Revised 12/20/00