1.3.15
USER CONFIGURATION
The User Configuration Mode allows the operator the opportunity to set custom default
settings. These custom default settings will be used each time the Accutorr Plus is
turned on. Once the User Configuration Mode is entered, the only way to exit this mode
is to turn off the Accutorr Plus using the ON/STANDBY key (19).
1. To enter the User Configuration Mode, press and hold the DEFAULTS key (14) while
turning the unit ON. Release after the third beep.
2. To select a User Configuration item number, press the ROOM/BED key (24) to
display the desired User Configuration Number in the ROOM and BED displays
(25 & 26). See table below for User Configuration Numbers. The current default
setting for that item displays.
3. Press the NIBP START key (38) to be able to change the default value. The default
setting flashes.
4. Press the Patient Info. Up or Down Arrow key (27 or 28) to change the default setting.
5. Press the START NIBP key (38) to enter the changed default setting.
6. Repeat step 2 for additional choices.
The following table list the functions that can be configured in the user configuration mode.
User
Function
Configuration
Number
1a
Clock Set
1b
Date Format
2
Reserved for future
use.
3
Text / Symbols
4
Patient Size
5a
Time Out, LEDs
and LCD Characters
when unit is
powered from AC
mains.
5b
Time Out, LEDs
and LCD
Characters when
unit is powered
from the internal
battery.
*1231 represents 12 months/31 days and 3112 represents 31 days/12 months.
Accutorr Plus Service Manual
Chapter 1, Operation
Description
Setting the date and time. See
section 1.3.13 for details on setting
the clock.
Set the format as M/D/Y (1231)*
or D/M/Y (3112)*
Set the description of which alarm
limit is being set, Hi and Lo or the
graphic
. Also change the
Interval of OFF to ——.
Set the default patient size to be
Adult, Pediatric and Neonate.
Set how long the numeric
information is displayed, when no
keys have been pressed, in the
LEDs and LCD before they are
blanked to conserve energy. The
choices are: 5, 15, 30 or 60
minutes. NOTE: The information
is not erased.
Set how long the numeric informa-
tion is displayed, when no keys
have been pressed, in the LEDs
and LCD before they are blanked
to conserve energy. The choices
are: 5, 15, 20 or 30 minutes.
NOTE: The information is not
erased.
Factory Default
D/M/Y (3112)*
The word "Hi" which will
then use Hi and Lo as
the indicators. OFF for
Interval.
Adult
15 minutes
5 minutes
Revised 02/15/00
1-49
Need help?