Datascope Accutorr Plus Service Manual page 101

Hide thumbs Also See for Accutorr Plus:
Table of Contents

Advertisement

Battery
Battery Type:
Sealed Lead Acid
Number of Batteries:
1
Battery Voltage:
12 VDC nominal
Battery Capacity:
2.3 Amp-Hour
Battery Run Times
Accutorr Plus NIBP
from full charge
– 5 Hour
with a new battery at
Accutorr Plus NIBP
25° C with 1 NIBP
with Trend Screen
measurement every 5
– 3 Hours
minutes and recorder
in use.
Accutorr Plus NIBP with
Trend Screen and SpO2:
Recharge Time:
8 hours max.,
in standby only
Cyclic Service Life:
150 cycles, 100% to
20% of capacity
400 cycles, 100% to
40% of capacity
Real Time Clock
Resolution:
1 minute
± 1 minute/week
Accuracy:
Display Format:
24 hours
Power:
The real time clock maintains the time and date when the
instrument is On or in the Standby mode, connected to AC
mains or running from internal battery for at least ten years
from original assembly. The real time clock will maintain time
and date even if the instrument's main battery is disconnected.
Accutorr Plus Service Manual
Chapter 3 - Specifications
Lithium-ion
1
10.8 VDC nominal
3.6 Amp-Hour
Accutorr Plus NIBP – 8 Hours
Accutorr Plus NIBP with with
Trend Screen – 8 Hours
Accutorr Plus NIBP with
Trend Screen and SpO2:
Datascope – 1.5 Hours
Nellcor – 2 Hours
Masimo – 2 Hours
2 hours max., in standby only
500 cycles, 100% to 30% of capacity
1000 cycles, 100% to 50% of capacity
Datascope – 3.5 Hours
Nellcor – 4.5 Hours
Masimo – 4.5 Hours
3-5
Revised 12/20/00

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Accutorr Plus and is the answer not in the manual?

Table of Contents