7.8 NIBP NORMAL OPERATION

1. Press the left arrow key on the "Cufflink" Simulator to highlight SELECT BP,
then press the ENT key. Set pressure test to 120/80 (90) and press the ENT key.
2. Attach the simulator "Cufflink" to the input pressure fitting of UUT.
3. Press the interval key of the keypad until a 1 is shown in the Timer/Temp./Interval
LED of the UUT. Press the Patient Size key to select Adult size indicator.
4. Press the Set Alarms key to enter the alarm setting mode. Use the Up and Down
arrow keys to set the Systolic Low alarm to 115 mmHg.
5. Press the Set Alarms key again and set the Diastolic Low Alarm to 95 mmHg.
6. Press the Start NIBP key. Verify the monitor takes a measurement and the Systolic
and the Diastolic low alarms are violated. The LED's for these parameters will blink
on and off and an alarm tone will sound for the violation. Press the Mute Key to
silence the alarms.
7. After the fifth measurement is complete, allow the cuff to pump up. Press Deflate
key. Verify cuff deflates.
8. Review the data that was stored in the database, by scrolling through the patient trend
information. Use the Up and Down Room and Bed keys or LCD display of the UUT
and verify that the Systolic and Diastolic information is available in the LED's for each
stored settings
9. Remove the cuff from the UUT and press the Start key. Confirm that the UUT pump
runs and that it displays a "-10" in the Systolic LED within 60 seconds and attempts
another inflation. Press the Deflate key to stop test.
7-20
Revised 03/25/02
Accutorr Plus Service Manual
Chapter 7 - Calibration