Incorrect placement can also reduce the acquired sensor signal, and therefore compromise
performance. Select an alternate site (toe) or use a FLEXISENSOR® SD if the sensor can
not be placed on the patient's finger correctly or if the fingernails interfere with the
acquisition of a reliable signal.
·
Use of the DATASENSOR is not recommended for long-term monitoring (4-6 hours). For
monitoring situations exceeding 4-6 hours, either reposition the DATASENSOR every
2-4 hours to a different site (finger/toe) or use a FLEXISENSOR
bandage.
·
Do not over-tighten the sensor bandages. Excessive pressure on the monitoring site can
affect SpO
also result in pressure necrosis and other skin damage.
·
Sensor configuration provides virtually uninterrupted monitoring during following
situations:
Electro-cautery Noise - Electro-cautery noise rejection is designed into the sensors.
Motion Artifact - The monitor's software adjusts the "averaging period" increasing it
during motion and reducing it during inactivity. This decreases the number of
monitoring interruptions and false alarms.
Weak Peripheral Pulses - The monitor's gain is automatically increased to track
pulses on patients with decreased peripheral perfusion.
Datascope SpO
Sensors
Approximate
Patient Weight
Where Used
Long or Short
Term Moni-
toring
Electro-Surgical
Interference
Suppression
(ESIS)
Reusable
Bandage Type
Part
#'s**
* <Non-adhesive bandages are recommended for premature infants to minimize prenatal skin damage.
** See Accessories, Chapter 5, for more detailed information.
*** Additional choices: 0060-00-0026-02 (10' sensor cable), 0020-00-0071-01
(3' sensor cable plus 7' extension cable).
1-36
Revised 02/15/00
readings and may reduce readings below true SpO
2
Sensor Selection Table
2
Large
Adult
Adult (LA)
(A)
>80kg/
0 - 90kg/
>176 lbs
66 - 198
lbs
Fingers,
Fingers,
Toes
Toes
Long &
Long &
Short
Short
Term
Term
Included
Included
Yes Up to
Yes Up to
20 Uses
20 Uses
Adhesive,
Adhesive,
Disposable
Disposable
Sensors 0998-00-0
0998-00-
076-06
0076-05
Bandages 0683-00-0
0683-00-
409-01
0409-02
®
SD with its appropriate
. Excessive pressure can
2
Pediatric
Infant
Adult
(P)
(I)
Ear (AE)
10 -
4.5 -
>40kg/
40kg/
10kg/
>88 lbs
22 - 88
10 - 22
lbs
lbs
Fingers,
Feet,
Adult
Toes
Palms,
Ear
Big Toes
Long &
Long &
Long &
Short
Short
Short
Term
Term
Term
Included
Included Included Included
Yes Up
Yes Up
Yes Up
to 20
to 20
to 20
Uses
Uses
Uses
Adhesive,
Non-
Adhesive Disposable
Disposable
Adhesive*
0998-00-
0998-00-
0998-00-
0076-04
0074-03
0074-05
0683-00-
0683-00-
N/A
0409-03
0415
Accutorr Plus Service Manual
Datasensor
40+ kg/
90+ lbs
Fingers, Toes
Short Term
Yes
6-Months
0600-00-0026-01
(3' sensor cable)***
N/A