1.3.5.3 How to Mute Alarms

When an NIBP alarm ex ists, press the MUTE key (30) (1 beep tone) to si lence the alarm
tone for 2 min utes. The alarm tone will re turn af ter the next measurement value that
violates the selected limits.
When an SpO
alarm ex ists, press the MUTE key (30) (1 beep tone) to si lence the alarm
2
tone for two min utes. The alarm tone will re turn af ter two min utes, un less the SpO
value changes and is within the alarm lim its. If dur ing that two min utes the mea sured
SpO
value changes to a value that is within the ac cept able range, and then re turns to
2
a value that is out side the set alarm limit, the alarm tone will re turn be fore the two
min utes elapse. Ex am ple (within 2 min utes): • SpO
• SpO
is mea sured at 89; the alarm tone sounds and the SpO
2
MUTE key is pressed. • SpO
dis play flashes. • SpO
is mea sured at 91; no alarm tone sounds and the dis play stops
2
flash ing. • SpO
is mea sured at 89; the alarm tone sounds and the SpO
2
Press and hold the MUTE key (30) (2 beep tones, approx. 3 sec onds) to per ma nently
si lence the alarm tone. The MUTE LED (31) il lu mi nated. The LEDs for the alarm ing
pa ram e ter will con tinue to flash. To re ac ti vate the alarm tone func tion, press the MUTE
key (30) again.
1.3.5.4 Alarms and Changing the Room Num ber and/or Bed Let ter
When chang ing the rooms and beds, the alarm set tings will change if the fi nal room/bed
dis played is a dif fer ent pa tient size than the orig i nal room/bed. When a new pa tient size is
de tected, the alarm set tings change to the de faults for the dif fer ent pa tient size. See sec tion
1.3.15 for in for ma tion on cus tom de faults.
The ta ble be low de scribes 6 mea sure ments in dif fer ent rooms/beds and dif fer ent pa tient
sizes, and the ef fect on the alarm set tings.
Mea sure ment Or der
Room/Bed
1
1/a
2
1/b
3
2/a
4
3/a
5
4/a
6
1/a
NOTE: The alarm set tings can be changed, if nec es sary, when chang ing the room/bed
and the pa tient size is the same.
PRE CAU TION: It is the us ers re spon si bil ity, when chang ing the room/bed, to as sure the pa tient
size and alarm set tings are as re quired.
Accutorr Plus Service Manual
Chapter 1, Operation
low alarm limit has been set to 90.
2
is mea sured at 88; there is no alarm tone, but the SpO
2
Pa tient Size
Adult
Adult
Pe di at ric
Adult
Adult
Adult
2
dis play flashes. • The
2
dis play flashes.
2
Alarm Set tings
Have been man u ally set.
Re main the same.
Changed to de faults for a
pe di at ric size pa tient.
Changed to de faults for an adult
size pa tient.
Re main the same.
Re main the same. If the alarm
set tings that were set from the
1st mea sure ment are re quired,
they need to be set again
man u ally.
Re vised 05/22/02