8.3 PREVENTIVE MAINTENANCE SCHEDULE

The following preventive maintenance steps are required for continued satisfactory
performance and safety of the Accutorr Plus and optional accessories. Inspections and
replacement of consumable supplies and accessories that are subject to normal wear, must
be accomplished at least as frequently as the inspection of the host unit. Read the
Warranty statement in the Operating Instructions for description of warranty conditions.
The suggested minimum maintenance schedule is based on normally expected wear and
tear of system and components. The intervals may be shortened at the user's discretion
when conditions of use may warrant. In case of internal instrument failure, component
replacement, and possible or actual instrument damage, it is strongly recommended that
complete performance and safety verification be conducted.

8.3.1 Mechanical and Physical Visual Inspection (One Year Interval)

Check the following items for wear and physical damage: Repair or replace as needed.

8.3.2 Electrical Safety and Performance Checks (One Year Interval)

8-2
Revised 2/15/00
Outer case of Accutorr Plus, Recorder and Temperature module.
Patient connected accessories such as blood pressure cuffs and hoses and related
connectors. Check the SpO
2
probes. Verify and replenish the disposable items as needed. (IE: probe covers,
recorder chart supply).
Power cords, external safety ground connections and communication cables.
Line power and battery power indicator lights functional in the proper modes.
Line power and safety grounds are intact and within specifications.
Non-invasive blood pressure section is functional within specifications.
SpO
functions are operational and within specification.
2
Temperature functions are operational and within specifications. Check the
temperature module battery voltage under load: replace if under 8.8 V. Use only
alkaline batteries. NOTE: Newer PTM modules do not contain a 9V battery.
Recorder is functional and documents available stored data.
Error codes, failure messages, end user reported anomalies are collected and
analyzed for reduced performance trends. Identify and correct causes as warranted.
Check recorder paper path and clean as required.
Perform ground resistance and chassis leakage tests.
sensor and cables. Check the temperature cables and
Accutorr Plus Service Manual
Chapter 8 - Preventive Maintenance