3.3 Physical Characteristics
Size (maximum):
Main Unit:
Recorder Module:
Predictive Module:
Weight:
3.4 Environmental Characteristics
Operating Temperature:
Accutorr Plus with or
without Recorder:
Infrared Therrmometer
Module:
Predictive Thermometer
Module:
Operating Humidity:
Shock and Vibration:
Shipping:
Storage Temperature:
Storage Humidity:
Operating Altitude:
3.5 Electrical Ratings
Voltage:
Current:
Frequency:
Power Consumption:
Accutorr Plus Service Manual
Chapter 3 - Specifications
19 cm(W) x 26.93 cm(H) x 20.83 cm (D)
7.5" (W) x 10.6" (H) x 8.2" (D)
5.33 cm(W) x 23 cm(H) x 11 cm (D)
2.1" (W) x 9" (H) x 4.25" (D)
5.7 cm(W) x 15.9 cm(H) x 11.8 cm (D)
2.25" (W) x 6.25" (H) x 4.63" (D)
<4.95 kg (11 pounds), depending on configuration.
10° C to 40° C, (50° F to 104° F)
18° C to 40° C, (65° F to 104° F
10° C to 32° C, (50° F to 90° F)
15 to 90% max, non-condensing.
Meets IEC 68-2-27, IEC 68-2-37.
Meets ISTA Test Procedure 1A (less than 100 lbs.)
-15° C to +40° C, +5° F to 104° F
10 to 95%, non-condensing.
1013 hPa to 782 hPa (0 to 7,000 ft.). for units with an LCD
1013 hPa to 697 hPa (0 to 10,000 ft.) for units
without an LCD
100 - 120 / 220 - 240 VAC
0.6 / 0.3 A
60 / 50 Hz
40 W, maximum
3-7
Revised 12/20/00