Typical Operation
Fresh Gas Failure Detection
During operation, the Apollo checks that the piston
cylinder unit has a sufficient level of fresh gas.
If the message "FGAS LOW OR LEAK" appears,
increase the fresh-gas flow; identify and repair leak.

Setting Vaporizer Concentration

Note:
Refer to the appropriate Instructions for Use
for the vaporizer being used. Vapor 2000 is
shown and described below.
For the Dräger Interlock 2 exclusion system:
1. Lock the unused vaporizer by moving the
selector lever completely towards it. For
example, to lock the left vaporizer, move the
lever to the left (1 in
2. With the handwheel set to "T" position on the
unlocked vaporizer, press the button and engage
the handwheel at "0" (2 in
seconds for the pressure to balance.
3. Press the button and turn the handwheel
counterclockwise to set the required anesthetic
gas concentration (3 in
For the Dräger Auto Exclusion system:
1. Close any open vaporizers.
2. With the handwheel set to "T" position, press the
button and engage the handwheel at "0". Wait
five seconds for the pressure to balance.
3. Press the button and turn the handwheel
counterclockwise to set the required anesthetic
gas concentration.

Setting Ventilation Mode

Set the ventilation mode as described in the
"Ventilation" section of this manual.
Apollo Operator's Manual
Figure
55.).
Figure
55.) Wait five
Figure
55.).
Chapter 6 - Operation Summary
Figure 55. Setting Vaporizer Concentration - Dräger
Interlock 2 Exclusion System
1
0
3
0
2
75