Ags Scavenger - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Chapter 13 - Specifications

AGS Scavenger

Technical Data of AGS Scavenger
Ambient conditions
During operation:
Temperature 59 °F to 86 °F
Atmospheric pressure
Rel. humidity 0 to 98%, without condensation
During storage:
Temperature –4 °F to 158 °F
Atmospheric pressure
Rel. humidity 0 to 98%, without condensation
Performance data
Effect on the anesthetic system/ventilator:
Increase in exhalation <0.5 cmH
resistance
<3.6 cmH
Generated vacuum pressure
Generated flow
Effect on the environment:
Gas loss,
dependent on fresh gas flow & ventilation parameters <25 mL/min typical)
Operating data
Operating capacity of scavenging flow
Setting with service flow indicator
Particle filter pore width
Particle filter resistance
Supply performance requirements
across a flow resistance of 20 cmH
across a flow resistance of 10 cmH
Dimensions (W x H x D)
Weight, without hoses
220
525 to 795 mmHg
375 to 825 mmHg
O at 30 L/min
2
O at 75 L/min
2
<1 Pa (0.01 cmH
<50 mL/min
30 to 50 L/min
65 to 75 L/min
0.1 mm
5.1 cmH
O at 50 L/min
2
O
2
O
2
4.7 in. x 13.8 in. x 7.9 in.
approx. 2.1 lbs.
O)
2
min. 25 L/min suction f
max. 50 L/min suction
Specifications