Chapter 11 - Cleaning and Maintenance

Removing the passive scavenger system (optional):

1. Remove the waste gas hose connected between
the hospital waste gas disposal system and the
bottom connection on the scavenger (1 in
143).
2. Remove the scavenger hose connected between
the scavenger and the scavenger connection on
the back of the anesthesia machine (2 in
143).
3. Remove the scavenger system from the
machine.
4. Prepare the parts for cleaning and disinfection.
Note:
For cleaning/disinfection instructions for the
passive scavenger system, refer to its
Instructions for Use.

Removing the suction system (optional)

1. Remove the suction bottle from the slide mount
(1 in
Figure
144).
2. Remove the suction regulator from the bracket (2
in
Figure
144).
3. Remove the suction system bracket from the
side rail on the side of the anesthesia machine.
4. If using disposable suction bottle, dispose of the
bottle with infectious waste.
5. If using reusable suction bottle, empty the bottle
and dispose of contents with infectious waste.
WARNING !
Always wear gloves when emptying suction
bottle.
Note:
For cleaning/disinfection instructions for the
reusable suction bottle, refer to its
Instructions for Use.
Note:
For cleaning/disinfection instructions for the
suction regulator, refer to its Instructions for
Use.
172
Figure 143. Passive Scavenger System
Figure
Figure
Figure 144. Suction System Bracket and Regulator
1
Apollo Cleaning and Disinfection
2
1
Apollo Operator's Manual
1
2
2