Specifications
Interfaces
2 serial interfaces: COM1, COM2
Protocol
Plug connector
Pin assignment
1
2
3
4
5
6
7
8
9
Shields
Settings
Dräger Base IV system (not sold in the U.S.)
SpO
2
1 typical delay of system alarms: 600 ms
CLIC Adapter Technical Data
Resistance
Compliance
Normal mode:
Bypass mode:
Leakage:
Apollo Operator's Manual
1
MEDIBUS
(COM 2 without real-time data)
9-pin sub-D, electrical isolation, 1.5 kV
NC, not connected
TX, transmit
RX, receive
DTR, data terminal ready
GND, ground
DSR, data set ready
RTS, request to send
CTS, clear to send
NC, not connected
DTR and DSR, as well as RTS and CTS are internally connected.
Hardware handshake is not supported.
1200 or 9600 Baud
even parity
8 data bits
1 stop bit
For connecting IV systems
For connecting an SpO
sensor
2
Reduced by approx. 1 mbar in the bypass mode, depending on the
breathing system of the anesthetic workstation
typically 1.3 mL/mbar
typically 0.4 mL/mbar
<5 mL/min
Chapter 13 - Specifications
219