Apollo Maintenance; Definitions; Maintenance Intervals - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Apollo Maintenance

Apollo Maintenance

Definitions

Inspection:
examination of actual condition
Service:
measures to maintain specified condition
Repair:
measures to restore specified condition
Maintenance: inspection, service, and repair, where
necessary
Preventive
Maintenance: maintenance measures at regular
intervals

Maintenance intervals

The workstation and its components must be cleaned
and disinfected before every maintenance operation,
especially before returning for repair.
Water trap
O
sensor
2
Flow sensors
AGS filter
Filter of the endotracheal aspiration
system
Nafion hose on patient gas module with
bacterial filter
Filter mat, patient gas module
Filter mat, power pack
Filter mat, housing cover
Dust filter, ventilator unit
Upper diaphragm of ventilator unit
Lower diaphragm of ventilator unit +
O-ring
O-rings, Vapor plug system
Pressure reducer mixer for piped medical
gas supply
Sintered filter, gas supply block
PEEP diaphragm, breathing system
Man. Spont. – Automatic reversing
diaphragm
Lead gel battery in UPS
Inspection and maintenance
Apollo Operator's Manual
Replace when dirty or when the message
CHECK SAMPLE LINE is displayed (if the sample line is free of blockages and not
kinked) and at least every four weeks.
Replacement of O
sensors is not necessary in conjunction with consumption-free
2
O
measurement.
2
Replace when calibration is no longer possible or if an error message is displayed.
Replace when blocked.
Replace every two weeks.
Must be replaced every two years by authorized representative of DrägerService.
Must be replaced once per year by authorized representative of DrägerService.
Must be replaced once per year by authorized representative of DrägerService.
Must be replaced every two years by authorized representative of DrägerService.
Must be replaced once per year by authorized representative of DrägerService.
Must be replaced after one year.
Must be replaced every three years by authorized representative of DrägerService.
Must be replaced once per year by authorized representative of DrägerService.
Must be replaced every six years by authorized representative of DrägerService.
Must be replaced every two years by authorized representative of DrägerService.
Must be replaced every two years by authorized representative of DrägerService.
Must be replaced every two years by authorized representative of DrägerService.
Must be replaced by authorized representative of DrägerService when the
message BATTERY LOW is displayed or every three years.
Must be disposed of in accordance with local waste disposal regulations.
Must be carried out once per every 6 months by authorized representative of
DrägerService.
Chapter 11 - Cleaning and Maintenance
185

Advertisement