Chapter 2 - System Components
Dräger Auto Exclusion 2-Vaporizer
Mount (Optional)
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When one of the two vaporizers is selected for use
(opened), the interlock mechanism within that
vaporizer's mounting system is activated
automatically, preventing the other vaporizer from
being used.
Note:
Only vaporizers labeled as "AUTO
EXCLUSION" vaporizers are compatible with
the Dräger Auto Exclusion 2-Vaporizer
Mount. See
Table 1
Vaporizer technical data.
When using a Desflurane vaporizer, it must be
plugged into the auxiliary power outlet located on the
side of the machine above the vaporizer mount.
Dräger Auto Exclusion 3-Vaporizer
Mount (Optional)
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When any one of the three vaporizers is selected for
use (opened), the interlock mechanism within that
vaporizer's mounting system is activated
automatically, preventing the other two vaporizers
from being used.
Note:
Only vaporizers labeled as "AUTO
EXCLUSION" vaporizers are compatible with
the Dräger Auto Exclusion 3-Vaporizer
Mount. See
Table 1
Vaporizer technical data.
When using a Desflurane vaporizer, it must be
plugged into the auxiliary power outlet located on the
side of the machine above the vaporizer mount.
Note:
The Desflurane vaporizer should be installed
in the far left position (1 in
Dräger Auto Exclusion 3-Vaporizer Mount in
order to have optimum viewing area of the
display screen.
24
for the Auto Exclusion
for the Auto Exclusion
Figure
5) with the
Vaporizer Exclusion Systems
Table 1: Dräger Auto Exclusion Vaporizer Technical
Data
≤ 10 L/min
Normal
Operating
Range
>10 ≤ 15 L/min Dräger auto exclusion
Extended
Operating
Range
Figure 5. Dräger Auto Exclusion 3-Vaporizer Mount
Dräger Vapor 2000 Instruction
for Use Manual's delivered
concentration accuracy values
apply.
vaporizer concentration output
accuracy may be reduced.
Apollo Operator's Manual