Apollo Cleaning And Disinfection; Disassembling Components; Removing The Sample Line; Removing The Water Trap - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Apollo Cleaning and Disinfection

Apollo Cleaning and Disinfection
This section of the manual provides complete
instructions for the disassembly and cleaning of the
Apollo anesthesia machine.
Note:
Set the Apollo to Standby before
disassembly.

Disassembling Components

Removing the sample line

Disconnect the sample line from the Y-piece
and the fitting on the water trap (see
132).
CAUTION !
Sample gas lines are single-use articles and must
not be disinfected.
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.

Removing the water trap

Pull the water trap straight out of its holder.
WARNING !
The water trap must be replaced after a service
life of four weeks.
The diaphragm could become brittle after an
extended period of use, allowing water and
bacteria to get into the measurement system.
The contamination affects the gas
measurement, or the gas measurement may fail
as a result.
CAUTION !
Do not allow any alcohol or cleaning agents/
disinfectants to come into contact with the inside of
the water trap.
Do not wash, flush or sterilize the water trap.
These substances can damage the diaphragm and
the measurement system.
Apollo Operator's Manual
Chapter 11 - Cleaning and Maintenance
Figure 132. Removing the Sample Line
Figure
Figure 133. Removing the Water Trap
167