Automatic Parameter Changes

Frequency Changes

If so configured, the inspiratory time (T
automatically changed when the frequency (Freq) is
changed in volume-controlled or pressure-controlled
ventilation modes (without synchronization), so that
the resultant ratio of inspiration to expiration (I:E)
remains constant.
To make a combined Freq/T
1. Press the >Freq< soft key on the Volume or
Pressure mode screen. The Freq key lights up
yellow, along with the value for T
that both values will change (1 in
2. Turn the confirm knob to adjust the Freq value.
The T
value is adjusted at the same time.
INSP
3. When the desired Freq value is displayed, press
the confirm knob to set the value. Both the Freq
and the T
keys turn green and the I:E ratio
INSP
remains constant.

PEEP Changes

If so configured, the inspiratory pressure (P
be automatically changed when the PEEP value is
changed in the pressure-controlled ventilation mode.
To make a combined PEEP/P
change:
1. Press the >PEEP< soft key on the Volume or
Pressure mode screen. The PEEP key lights up
yellow, along with the value for P
that both values will change (1 in
2. Turn the confirm knob to adjust the PEEP value.
The P
value is adjusted at the same time.
INSP
3. When the desired PEEP value is displayed,
press the confirm knob to set the value. Both the
PEEP and the P
INSP
In addition, if so configured, the PAW low alarm limit
will be automatically changed when the PEEP value
is changed.
Apollo Operator's Manual
) will be
INSP
parameter change:
INSP
, to indicate
INSP
Figure
73).
) will
INSP
parameter
INSP
, to indicate
INSP
Figure
74).
keys turn green.
Figure 73. Automatic Freq/T
1
Figure 74. Automatic PEEP/P
1
Chapter 7 - Ventilation
Change
INSP