Reassembling Components

Reassembling the AGS scavenger system (optional)

1. Install the particle filter and tighten the union nut
(1 in
Figure
152).
2. Reinstall the flow tube (2 in
scale facing the front of the machine, and tighten
the union nut (3 in
Figure
3. Reinstall the buffer volume container into the
scavenger body.

Connecting the AGS Scavenger System (optional)

The AGS scavenger system is used with vacuum
waste gas disposal systems.
WARNING !
The scavenger hoses must not be pinched,
kinked, or blocked in any manner.
1. Install the AGS scavenger on the machine by
sliding its bracket onto the two shoulder screws
on the side of the machine.
2. Connect one end of the transfer hose to the
fitting on the scavenger (1 in
3. Connect the other end of the transfer hose to the
scavenger connection on the back of the
anesthesia machine (2 in
4. Connect the waste gas vacuum hose to the
output connection on the scavenger (3 in
153).
5. Connect the other end of the vacuum hose to the
hospital waste gas disposal system.
Note:
Activate hospital vacuum system before
using scavenger system.
Note:
During use, the float indicator in the flow
indicator should stay between the upper and
lower marks. If necessary, regulate flow
using the flow adjustment valve (4 in
153).
WARNING !
Do not block the openings in the AGS
scavenger system. Blocking the openings could
result in a shortage of fresh gas in the breathing
system.
Apollo Operator's Manual
Figure
152) with the
152).
Figure
153).
Figure
153).
Figure
Figure
Chapter 11 - Cleaning and Maintenance
Figure 152. Reassembling the AGS Scavenger