Chapter 12 - Troubleshooting
Priority
Message
Advisory FAN FAIL
Caution
FG FLOW TOO HIGH
Caution
FG LOW OR LEAK
2
Warning
Advisory GAS SENSOR FAIL
Warning/
GAS MIXER FAIL*
Advisory
Warning/
GAS + VENT FAIL.*
Advisory
Warning HIGH AIRWAY PRESSURE
Caution
INCORRECT FG FLOW
Caution
INSP CO
HIGH
2
Advisory INSP. FLOW SENSOR FAIL
1. Defective fans in combination with an internal leakage may lead to elevated O
Risk of fire!
2. The priority of the alarm message depends on the severity of the fresh gas shortage.
3. If the alarms "VT NOT ACHIEVED" or "PINSP NOT ACHIEVED" occur additionally, a fresh gas deficiency in the system is remedied by
entraining ambient air. Monitor concentrations carefully as the breathing gas will be diluted.
4. Displayed in Standby mode only.
5. At high ventilation frequencies, the measured value cannot keep up completely with the gas concentration for reasons due to the system.
*
Alarm can be downgraded to specified lower priority or canceled.
200
Cause
Fan for evacuating gases inside the
device is defective.
Total fresh-gas flow is above 19 L/min.
3
fresh gas setting too low.
Leak.
4
Complete gas measurement system
failure.
Fresh gas measurement is probably
inaccurate or failed.
Switch-over to auxiliary CGO may have
failed.
Ventilator failed.
Fresh gas measurement is probably
inaccurate or failed.
Switch-over to auxiliary CGO may have
failed.
High alarm limit for the airway pressure
has been exceeded.
Ventilation hose kinked.
Stenosis.
Ventilation settings are not correct.
Reduced accuracy of fresh-gas flow
measurement.
Soda lime in circle system exhausted.
Leak or fault in breathing system.
High ventilation frequencies
Dead space ventilation.
Inspiratory flow sensor is faulty.
Alarm – Cause – Remedy
Remedy
Workstation must be switched off as
quickly as possible!
Call DrägerService.
Reduce fresh-gas flow.
Check vaporizer setting.
Increase fresh-gas flow.
Repair leak.
Use external gas measuring system.
Call DrägerService.
Use only oxygen as fresh gas and check
total flow meter.
Call DrägerService.
Patient must immediately be ventilated
manually!
Use only oxygen as fresh gas and check
total flow meter.
Call DrägerService.
Check hose system and tube.
Correct the ventilation settings.
Reduce fresh-gas flow for each gas to
below 12 L/min.
Check total flow meter.
Call DrägerService.
Increase fresh-gas flow.
Replace soda lime.
Replace breathing system.
5
.
Adjust alarm limits if necessary.
Check ventilation settings.
Replace flow sensor (see page 170).
concentrations inside the workstation.
2
1
Apollo Operator's Manual