Fresh Gas Data; Auxiliary O2 Flowmeter - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Specifications

Fresh Gas Data

Settings:
O
concentration
2
Fresh-gas flow, dosing
Fresh-gas flow, electronic measuring
Accuracy
Resolution
Fresh-gas flow, total flowmeter
Accuracy
Resolution
flush
O
2
Auxiliary O
Flowmeter
2
Connection
Fresh-gas flow
Accuracy
Resolution
Apollo Operator's Manual
21 to 100 Vol.%
0 to at least 12 L/min per gas (O
0 to 12 L/min volumetric flow per gas (O
±10% or 0.12 L/min, whichever is greater
0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from >0.2 to 0.5 L/min)
0.05 L/min (from >0.5 to 1.0 L/min)
0.1 L/min (from >1 to 12 L/min)
0 to 10 L/min
±10% full scale for 50% O
2
calibrated at 5 L/min
0.5 L/min (from 0.5 to 2 L/min)
1.0 L/min (from 2 to 10 L/min)
>35 L/min
Staged connector for use with different hose diameters
0 to 10 L/min
± 5% of full scale
0.5 L/min
Chapter 13 - Specifications
, N
O, Air)
2
2
, N
O, Air)
2
2
/ 50% N
O
2
211