Checking the Workstation According to the Check List
13. If all checks in the check list are completed
successfully, press the confirm knob or the
>Start Self Test< key on the check list screen (1
in
Figure
44) to begin the Apollo automated self
test.
The automatic self test lasts approximately three
minutes. The bar graph at the top of the Self Test
screen shows the progress of the test
45).
The tests that are performed are listed on the
screen. A clock symbol >
small circle next to the component that is
currently being tested. As each component test
is finished, the clock symbol is replaced by a
color code that indicates the result of the test.
Green
Test completed successfully.
Yellow
Workstation can be used with restrictions.
The error that was found during the test is
posted in the window at the bottom of the
screen with information on how to
remedy the problem
The user can either press the confirm
knob to repeat the test, or can confirm the
error by pressing the >Accept< key
displayed on the screen. If the error is
accepted, the workstation starts
operation without this function.
Red
Operation of the workstation is
impossible or not permitted. The error
must be remedied and the test must be
repeated.
The self-test can no longer be cancelled
at this point.
CAUTION !
Functions highlighted in yellow during the self test
may not meet specifications and therefore merit
further investigation and resolution by a properly
trained technician.
Apollo Operator's Manual
(Figure
u
< is displayed in the
(Figure
46).
Chapter 5 - Preuse Checkout
Figure 44. Check List Screen
Figure 45. Self Test Screen
1
63