Spo2 Alarms On/Off (Optional); Bypass Mode - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Chapter 9 - Alarms
SpO
Alarms On/Off (optional)
2
The SpO
alarms can be disabled/enabled while in
2
any ventilation mode via the configuration screen:
1. Press the >Config< soft key (1 in
the standard or data screen.
The Volumes/Alarms configuration screen is
displayed (refer to
Figure
2. Select and confirm the "Alarms On/Off" column
using the confirm knob.
3. Select and confirm the "SpO
confirm knob.
4. Select and confirm "On" or "Off" using the
confirm knob.

Bypass Mode

The bypass mode permits patient monitoring without
unnecessary alarms during extra-corporeal
oxygenation of the patient by a heart lung machine.
All gas concentrations are measured independently
of the breathing phase in bypass mode. It can be
used in all active ventilation modes.
When bypass mode is activated, all CO
pressure apnea alarms, and SpO
are automatically disabled.
To enable/disable bypass mode:
1. Press the >Config< soft key (1 in
the standard or data screen.
The Volumes/Alarms configuration screen is
displayed (refer to
Figure
2. Select and confirm the "Alarms On/Off" column
using the confirm knob.
3. Select and confirm the "Bypass Mode" option
using the confirm knob.
4. Select and confirm "On" or "Off" using the
confirm knob.
The indication "Bypass Mode" is displayed in the
status field when the mode is active (see
Bypass mode remains activated when changing
ventilation modes; it is deactivated when the
machine is set to Standby.
Deactivating bypass mode immediately enables all
CO
and pressure apnea alarms, but SpO
2
measurement and alarms are enabled only when a
pulse signal is detected.
140
Disabling/Enabling Alarms Globally During Operation
Figure
105) in
106).
" option using the
2
alarms,
2
monitoring alarms
2
Figure
105) in
107).
Figure
107).
2
Figure 106. Volumes/Alarms Configuration Page
Figure 107. Bypass Mode Indication
Bypass Mode Indication