Alarm – Cause – Remedy
Priority
Message
Caution
NEGATIVE PRESSURE
Warning
NO AIR SUPPLY*
/
Advisory
Warning
NO FRESH GAS
/Advisory
Warning
NO N
O SUPPLY*
2
/Advisory
Warning NO O
SUPPLY
2
Warning NO SPO
2
PULSE
Advisory/
O2 CYLIND. CONNECT.?*
1. Disconnect the connector from the wall supply in the event of a prolonged gas failure.
Otherwise the central gas supply may become contaminated!
*
Alarm can be downgraded to specified lower priority or canceled.
2. This alarm can be deactivated by trained service personnel at customer request.
Apollo Operator's Manual
Cause
Insufficient supply of fresh gas.
Bronchial aspiration during ventilation.
Negative pressure due to fault in
ventilator.
Anesthetic gas scavenging system faulty. Call DrägerService.
Compressed air supply has failed.
Pipeline supply hose not connected or
kinked.
Optional air cylinder is empty or closed.
Compressed air compressor has failed.
2
*
No fresh-gas flow set.
N
O supply has failed.
2
Pipeline supply hose not connected or
kinked.
N
O cylinder empty or closed.
2
O
supply has failed.
2
Pipeline supply hose not connected or
kinked.
O
cylinder empty or closed.
2
No pulse signal detected with the SpO
measurement for approx. 10 seconds.
NiBP measurement on the same arm.
Pressure sensor of backup cylinder not
connected.
Chapter 12 - Troubleshooting
Remedy
Set adequate fresh-gas flow on
anesthesia machine. Flush system if
necessary.
Check bronchial aspiration.
Check that upper diaphragm has been
fitted correctly.
Call DrägerService.
Open optional backup air cylinder.
Check piped medical air supply.
Check connection to piped medical air
supply.
Connect a full air cylinder or open the
cylinder valve.
Check compressor.
Open flow control valves and adjust flow.
Open N
O backup cylinder.
2
Check central gas supply.
Check connection to central supply.
Connect a full N
O cylinder or open the
2
cylinder valve.
Open O
backup cylinder.
2
Check central supply.
Check connection to central supply.
Connect a full O
cylinder or open the
2
cylinder valve.
Check patient's condition!
2
Check application of the SpO
Measure blood pressure on other arm.
Check pressure sensor connection.
1
1
)
1
sensor.
2
203