Installing The Breathing System; Filling And Installing The Absorber - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Chapter 11 - Cleaning and Maintenance

Installing the breathing system

Carefully seat the breathing system onto the
ventilator module, and tighten the three sealing
screws (1 in
Figure
cover.

Filling and installing the absorber

1. Push the insert fully into the absorber canister (2
in
Figure
149).
2. Fill the absorber canister with fresh soda lime up
to the MAX mark.
WARNING !
Absorbent is caustic and is a strong eye, skin,
and respiratory tract irritant. Use care in
handling the absorbent to avoid spills.
CAUTION !
It is recommended that Drägersorb 800 Plus or
Drägersorb FREE be used. Do not use powdered
soda lime, as a higher dust load may impair
functionality of the Apollo workstation.
3. Fit the canister into position below the breathing
system, and turn counterclockwise as far as
possible.
4. Slowly push in the ventilator module until it
engages.
180
148) on the ventilator
Reassembling Components
Figure 148. Installing the Breathing System
Figure 149. Installing the Absorber Canister
2
1
1
1
Apollo Operator's Manual