Fan Failed - Agilent Technologies G1313A Reference Manual

1100 series standard, micro and preparative autosamplers
Table of Contents

Advertisement

Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Fan Failed

Fan Failed
The cooling fan in the autosampler has failed.
The hall sensor on the fan shaft is used by the ASM board to monitor the fan
speed. If the fan speed falls below 2 revolutions/second for longer than
5 seconds, the error message is generated.
• Fan cable disconnected.
• Defective fan.
• Defective ASM board.
❏ Ensure the fan is connected correctly.
❏ Exchange fan.
❏ Exchange the ASM board.