Error Statuses and Response Messages
During operation, error statuses are indicated with sound and vibration signals.
Signal
1 × long beep,
1 × vibration
1 × long beep,
5 × vibration
2 × beep (low, low)
1 × vibration
Response Messages on Operating States
For response messages on the operating states, 4 different sound sequences and short vibrations in the
DynamicArm are used.
Sound sequence
INFORMATION
It is possible that only individual or several components can no longer be controlled.
60 | Ottobock
Error
Critical error
(e.g. a sensor is not
operational)
Severe error
(e.g. temperature of
the lifting motor too
high)
Malfunction
(e.g. incompatible
component
connected)
Event
Low, low,
DynamicArm has been
high,
turned on.
low
Very high,
high, low, very
DynamicArm has been
low
turned on
Battery charger was
connected during
6 x low
operation
Prosthetic component
Low, high
is being controlled
Annex
Error correction
Turn the DynamicArm off and on again or
connect the battery charger while the
DynamicArm is switched on.
Contact the Ottobock Myo-Service if the
error has not been resolved.
Prosthetic component
DynamicArm
Function
Present
No function
No function
Function of the prosthetic component in
question is not possible
Electric wrist
Hand/gripper
rotator
Function
Function
Present
Present
Function
Function
Present
Present
No function
No function
DynamicArm