2
DynamicArm
3
2
Wiring the DynamicArm with the Electric Wrist
Rotator 10S17
INFORMATION
The Electric Wrist Rotator 10S17 has numbers on
the plastic housing near the contacts. These numbers
will help when attaching the cables!
NOTICE
Combination with MyoRotronic 13E205. The Myo-
Rotronic 13E205 cannot be combined with the Dy-
22
namicArm.
The Myo R otronic 13E205 will be destroyed as soon
as the DynamicArm is operated!
Attach the motor cable to one of the two no. 3 contacts.
Attach two-pole connector to contact no. 3.
22a
Attach three-pole connectors to contacts no. 2. One
of the two three-pole connectors has a colour dot on
it (arrow).
INFORMATION
If the two three-pole connectors are attached the wrong
way round to the no. 2 contacts, the System Electric
Hand or System Electric Greifer will not function!
23
Next, mount the Electric Wrist Rotator 10S17 in the
lamination ring. See Fig. 24 and Fig. 25.