Serial Commands
B-2
MIE ADR-1500 Instruction Manual
If the units are scattering coefficient (1/Mm), the analog output menu
choices = {0, 1, 2, 3, 4, 5}
0 = DISABLED
1 = 10/Mm
2 = 100/Mm
3 = 1000/Mm
4 = 10000/Mm
5 = 100000/Mm
The response is ANALOGOUT [menu] "menu entry" [factor to be
applied to output voltage to yield output units (ie, mg/m
AUTOSTART [min] [hour] [day] [month] [status]
Returns, or programs, the automatic start feature parameters.
min and hour program the start time
day and month program the start day
status = {"on", "off"} The automatic start is enabled if "on" is selected.
BACKLIGHT [Status]
Returns, or programs, the backlight status.
Status = {"enabled" or "on", "disabled" or "off"}
BATTERY [battery voltage] [alkaline charge] [NiMH charge]
Returns the battery voltage with an estimation of charge time left in
percentage if the batteries are Alkaline or NiMH.
CALFACTOR [float value]
Returns, or programs, the user calibration adjustment factor. The value can
range from 0.001 to 10.000.
CONTRAST [integer value]
Returns, or programs, the LCD screen contrast setting.
Integer value = {1, 2, 3, ..., 255; Hint: display is only visible between 0 and
40}
DATE [day] [month] [year]
Returns, or programs, the date in the RTC.
Day, month and year are all integer values.
DISPLAY (or DISP, or D, or V 5) [line 1 (16 characters)] [line 2 (16
characters)] [state #]
Return the text currently displayed on the LCD. It comes in two quoted
strings for the 2 lines of the display. It is followed by a state-number. Note
that response is prefixed with "V 5"; this can be ignored.
3
or 1/Mm)]